Computational Fluid Dynamics Assessment Associated with Transcatheter Heart Valve Prostheses: A Position Paper of the ISO Working Group

Author(s)

Wei, Z. A., Sonntag, S. J., Toma, M., Singh-Gryzbon, S., & Sun, W.

Title

Computational Fluid Dynamics Assessment Associated with Transcatheter Heart Valve Prostheses: A Position Paper of the ISO Working Group

Date

2018

Publisher

Springer Science and Business Media LLC

Subject

Computational fluid dynamics
Fluid-structure interaction
Prosthetic heart valves
Thrombus assessment

Language

English

Abstract

The governing international standard for the development of prosthetic heart valves is International Organization for Standardization (ISO) 5840. This standard requires the assessment of the thrombus potential of transcatheter heart valve substitutes using an integrated thrombus evaluation. Besides experimental flow field assessment and ex vivo flow testing, computational fluid dynamics is a critical component of this integrated approach. This position paper is intended to provide and discuss best practices for the setup of a computational model, numerical solving, post-processing, data evaluation and reporting, as it relates to transcatheter heart valve substitutes. This paper is not intended to be a review of current computational technology; instead, it represents the position of the ISO working group consisting of experts from academia and industry with regards to considerations for computational fluid dynamic assessment of transcatheter heart valve substitutes.

Source

Cardiovascular Engineering and Technology, Volume 9, Issue 3, April 2018, pages 289-299

Rights

Copyright © 2018, Biomedical Engineering Society

Format

PDF

Type

Text

Bibliographic Citation

Wei, Z. A., Sonntag, S. J., Toma, M., Singh-Gryzbon, S., & Sun, W. (2018). Computational Fluid Dynamics Assessment Associated with Transcatheter Heart Valve Prostheses: A Position Paper of the ISO Working Group. In Cardiovascular Engineering and Technology (Vol. 9, Issue 3, pp. 289–299). Springer Science and Business Media LLC. https://doi.org/10.1007/s13239-018-0349-y

Files

manuscript nihms961540.pdf

Citation

Wei, Z. A., Sonntag, S. J., Toma, M., Singh-Gryzbon, S., & Sun, W., Computational Fluid Dynamics Assessment Associated with Transcatheter Heart Valve Prostheses: A Position Paper of the ISO Working Group. Cardiovascular Engineering and Technology, Volume 9, Issue 3, April 2018, pages 289-299, New York Tech Institutional Repository, accessed April 28, 2024, https://repository.nyitlibrary.org/items/show/3727

Position: 908 (4 views)